run_isotropic_scanner follows the main-scan reconstruction in
main_script_7TTerraVE12U.m: load_siemens_lowres, then per-physical-coil
bmMathilda, and sqrt(mean(abs(coilImage).^2)). No calibration or ROVir is
required. Coil images are combined incrementally and are not saved.
addpath('/Users/cag/forclone/monalisa_tutorial_ui');
plan=run_isotropic_scanner('/Volumes/Extreme SSD/261110');Arguments are (datasetFolder,outDir,nShotOff,scanSet). Defaults are a new timestamped
folder under <dataset>/recon_results/isotropic_scanner, and 50 discarded
initial shots, matching the 7T main-scan script. Existing output folders are
refused. All five raw/sequence pairings are checked before reconstruction.
For the six scans selected from the 13 November dataset:
plan=run_isotropic_scanner('/Volumes/Extreme SSD/261113',[],50,'seq702_705to709');This selects only seq702, seq706, seq707, seq705, seq708 and seq709, in that acquisition order, using each sequence's unique matching raw file. Settings remain 240³, 1-mm output spacing, 50 excluded shots, physical-coil xRMS and verified max-normalized float32 NIfTI. Each sidecar retains the sequence-file flip angle and separately records the user's ×2 LIBRE reporting convention; the reported angle is not a measurement of local excitation.
For the seq700 normal-shimming GRE/LIBRE pair, use the same reconstruction:
plan=run_isotropic_scanner('/Volumes/Extreme SSD/261110',[],50,'seq700_norm_shimming');
% The linear-shimming pair uses the same settings and a separate new batch:
plan=run_isotropic_scanner('/Volumes/Extreme SSD/261110',[],50,'seq700_linear_shimming');This selects only meas_*seq700_gre_norm_shimming.dat and
meas_*seq700_libre_norm_shimming.dat, verifies each against its matching
GRE/LIBRE sequence, and saves separate labelled folders. Each source has 199
shots; discarding 50 initial shots and each SI leaves 149 × 43 = 6,407 imaging
spokes. The matrix, FOV, kernel, coil combination and verified compact export
are identical to the seq701–705 path. Original sequence flip-angle definitions
are retained in reconstruction metadata.
seq700_linear_shimming selects only the two *_linear_shimming.dat files,
with the same GRE/LIBRE sequence matching and output checks. Its separate
labels prevent mixing normal- and linear-shimming outputs.
The reconstruction grid is 240 × 240 × 240 across the declared 240-mm cubic
FOV: 1-mm isotropic output voxels. Acquisition res_mm remains 0.5 mm in the
source metadata. Each scan has 500 × 44 readouts, including one SI per shot;
after exclusions, 450 × 43 = 19,350 imaging spokes remain.
The existing isotropic loader uses Pulseq ROTATIONS through bmTraj, scales
the maximum trajectory radius to N/(2*FoV), computes spherical Voronoi DCF
after shot/SI exclusion, then selects samples inside the 240³ Cartesian
k-space grid using bmLowRes. Gridding uses the 7T defaults: Gaussian window
3, parameters [0.61 10], normalized gridding, MATLAB IFFT and deapodization.
No added phase or gradient-delay correction is applied.
Each seq701–seq705 folder saves xrms_normalized.nii.gz,
reconstruction.json, reconstruction.log, and xrms_preview.png.
The existing verified exporter is shared with the anisotropic pipeline.
Normalization divides each volume by its own maximum, without clipping.
The JSON divisor restores the original magnitude scale at float32 precision.
NIfTI read-back must match exactly, with relative restoration error <1e-7,
before deleting the temporary magnitude MAT. No full coil-image or encoding
MAT is retained. Original raw/sequence files are preserved.
Coordinates are the nominal centred sequence grid, with 1-mm spacing and
first voxel centre (−120, −120, −120) mm; patient anatomical orientation has
not been established. The export performs no axis permutation or resampling.
batch_plan.mat records completion per sequence and for the batch.
/Volumes/Extreme SSD/261110/recon_results/isotropic_scanner/run_20261111_174401_195
All five scans completed with the defaults above. The five compressed NIfTIs total 226.73 MiB (45.29–45.45 MiB each). MATLAB verified exact normalized float32 read-back for every volume; maximum relative intensity-restoration error was 2.59e-8. An independent Python standard-library check verified all five NIfTI headers, 240³ payload sizes, 1-mm spacing, float32 datatype, MD5 checksums and gzip integrity. No temporary magnitude MAT or uncompressed NIfTI remains in the scan output folders. Raw and sequence files were unchanged by reconstruction. These checks establish numerical/export integrity, not image-quality or patient-orientation validation.
/Volumes/Extreme SSD/261110/recon_results/isotropic_scanner/run_seq700_norm_shimming_20261112_094850_901
Both GRE and LIBRE completed at 240³ with 1-mm spacing, 50 discarded initial
shots, 6,407 retained imaging spokes and the same [0.61 10] Gaussian kernel.
Separate seq700_gre_norm_shimming and seq700_libre_norm_shimming folders
contain verified normalized float32 NIfTI, reversible scaling JSON, logs and
previews. NIfTI sizes are 46.53 and 46.61 MiB respectively. Exact MATLAB
read-back passed; maximum relative intensity-restoration error was 2.43e-8.
Independent checks passed for MD5, gzip integrity, header geometry/datatype,
payload size and expected acquisition/method metadata. Temporary MAT and
uncompressed NIfTI files were removed after verification. Local verification:
results/seq700_norm_shimming_20261112_verification.json.
Both previews depict a spherical phantom, not the orbital subject. These scans can assess phantom excitation/shimming behaviour but are not direct anatomical controls for the lens-region finding.
/Volumes/Extreme SSD/261110/recon_results/isotropic_scanner/run_seq700_linear_shimming_20261112_100256_121
Both GRE and LIBRE completed with the same matrix, spacing, exclusion, kernel,
coil combination and export method as the normal-shimming pair. Verified
float32 normalized NIfTI sizes are 46.69 MiB (GRE) and 46.32 MiB (LIBRE).
Maximum relative intensity-restoration error was 2.37e-8. Exact MATLAB
read-back and independent NIfTI/header/payload/MD5/gzip checks passed. The
verification also matched the normal-shimming grid, kernel, retained spokes,
sample count, sequence hash, loader, trajectory function and gridding MEX.
Temporary magnitude MAT and uncompressed NIfTI files were removed only after
export verification. Local audit:
results/seq700_linear_shimming_20261112_verification.json.
Both previews show the spherical phantom. The LIBRE linear-shimming preview contains a prominent dark signal-loss band in the XY/XZ projections; its physical cause has not been established by these reconstruction checks.
The subsequent comparison is saved under
results/seq700_shimming_comparison_20261112/. It restores the JSON intensity
divisors, checks identical processing within each GRE/LIBRE pair, and compares
native z=0/y=0 slices plus central-volume linear/normal ratio profiles.
LIBRE profile minima occur near x=−39 and +42 mm (8.1% and 6.2% of normal
signal); GRE retains 98.6% and 89.8% at those same profile positions.
Raw headers show X-gradient offset 273→3759 in scanner units, unchanged
Y/Z offsets and listed higher-order shim currents, and centre frequency
297191177→297191221 Hz. Per-sequence reference/RF amplitudes are unchanged.
Pulseq mr.simRf on the actual exported RF pulses predicts LIBRE excitation
nulls near −940/+945 Hz. A Fourier small-tip calculation independently agrees
within 2 Hz. The 530-µs subpulse-centre spacing gives the simple cancellation
scale 1/(2×530 µs)≈943 Hz. Together these results strongly support off-resonance
LIBRE excitation cancellation under the altered X offset. No measured B0/B1
map was available; exact voxelwise frequency assignment and the full
steady-state intensity pattern remain unverified. This phantom result does
not establish the cause of the orbital lens-region crescent.
/Volumes/Extreme SSD/261113/recon_results/isotropic_scanner/run_seq702_705to709_20261113_155007_367
All six highlighted scans completed in acquisition order: seq702, seq706,
seq707, seq705, seq708 and seq709. The unselected seq720–723 scans were not
included. Each output is 240³ physical-coil xRMS, with 1-mm spacing, 50
discarded initial shots, 19,350 retained imaging spokes, and Gaussian
window 3 / parameters [0.61 10].
| Scan | TR (ms) | TE (ms) | Reported LIBRE FA | NIfTI size (MiB) |
|---|---|---|---|---|
| 702 | 6.7 | 2.875 | 12° | 45.26 |
| 706 | 13 | 2.875 | 12° | 45.90 |
| 707 | 13 | 4 | 12° | 45.34 |
| 705 | 6.7 | 2.875 | 20° | 45.41 |
| 708 | 13 | 2.875 | 20° | 45.31 |
| 709 | 13 | 4 | 20° | 45.30 |
The reported flip angle follows the user's ×2 LIBRE convention; the original sequence-file values remain separately recorded in every JSON. All six max-normalized float32 compressed NIfTIs passed exact MATLAB read-back and independent source-pairing, header/affine/datatype/payload, finite normalized values, metadata, MD5 and gzip-integrity checks. Source sequence hashes and raw sizes also matched. The six NIfTIs total 272.52 MiB. Temporary magnitude MAT and uncompressed NIfTI files were removed after verification.
Audit and log: results/selected_261113/verification.json,
results/selected_261113/matlab_batch.log, and completed.mat. The independent
verifier is results/selected_261113/verify_exports.py. These checks establish
numerical/export integrity, not image quality or patient orientation.