Current handover: HANDOVER_ANISOTROPIC_RECON_20261111.md.
This pipeline reads scanner Siemens .dat and matching Pulseq .seq; it does not run Koma.
addpath('/Users/cag/forclone/monalisa_tutorial_ui');
dataset='/Volumes/Extreme SSD/261110';
sigmas=[.61 .32 .20 .10];
% Read-only plan (no folders, no raw-signal loading):
plan=run_anisotropic_scanner(dataset,sigmas,20,[],false,true);
% Recompute only when intended, into a new timestamped batch:
% plan=run_anisotropic_scanner(dataset,sigmas,20);The independent anisotropic cell in main_script_7TTerraVE12U.m uses the same four widths. New runs save verified max-normalized float32 NIfTI plus metadata/logs/previews and one shared encoding MAT per sequence. The transient full physical-coil MAT is deleted only after NIfTI/encoding verification. The low-level recon_anisotropic_xrms still saves its full MAT if invoked directly.
/Volumes/Extreme SSD/261110/recon_results/anisotropic_scanner/retained_20261111
Each seq712_sigma0p61, seq712_sigma0p32, seq712_sigma0p2, seq712_sigma0p1 directory contains xrms_normalized.nii.gz, reconstruction.json, and the scanner validation/log/preview. The same four directories exist for seq713 and seq714. Shared encoding/seq71*.mat contains t, ve, ADC times, metadata and coordinate vectors, not coil images.
Load without losing the original comparison scale:
addpath('/Users/cag/forclone/monalisa_tutorial_ui/pipeline_utils');
r=load_anisotropic_nifti('/Volumes/Extreme SSD/261110/recon_results/anisotropic_scanner/retained_20261111/seq712_sigma0p32');
% r.xrms is restored to its original magnitude scale (float32 export precision).
% Pass false as argument 2 to read the normalized values instead.Normalization is single(double(xrms)/max(double(xrms(:)))); no clipping, interpolation or axis permutation. The JSON divisor reverses the amplitude scaling. Shared-scale comparisons must restore that divisor. NIfTI sform code 2 denotes the nominal sequence grid, not validated patient anatomy.
The scientific reconstruction remains physical Pulseq ROTATIONS, cycles/mm, 240 × 240 × 180 mm FOV, 2-mm isotropic nominal voxels, 20 discarded initial shots, 7,697 retained imaging spokes, spherical Voronoi DCF, Gaussian window 3/parameters [sigma 10], and physical-coil xRMS. No added phase or gradient-delay correction.
Only 0.61/0.32/0.20/0.10 remain. Widths 0.50/0.40/0.27/0.15 and full coil-image MATs were explicitly removed. Old comparison figures/metrics remain as history in comparisons; new orbital_retained_zm24.* and zm20.* show the four retained widths on one restored raw intensity scale. Historical source scripts are under comparisons/historical_scripts and refer to pre-cleanup experiments; use the current entry point for new scanner work.
Static phantom diagnostics are separate at /Volumes/Extreme SSD/261110/recon_results/water_phantom_diagnostics. Their active scripts use the retained shared encoding. The cleanup audit is in ../output_review_20261111/cleanup_completed/.